Weight |
4 lbs |
Dimensions |
15 × 10 × 8 in |
Thermometry Number of Inputs |
1
|
Measurement Type |
4‐lead single polarity
|
Excitation |
Constant current, 10μA or 1mA (0.01%)
A/D Resolution: 24 bit
|
Input Accuracy |
Sensor Dependent
|
Max Measurement Rate |
5 rdg/s
|
Supported Sensors |
Silicon diodes
RTD’s
Gallium Arsenide (25K & above) Ru0 1KΩ, Ru0 100KΩ, Pt 100Ω, Pt 1000Ω
Cernox
Carbon-Glass (4K & above)
|
Standard Curves |
Si410, Si430, Si435, DT470, DT670, CTI, PT100, PT1000, RO600, RO105
|
User Curve |
One 200‐point user curve in non‐volatile memory
|
Input Connector |
DB15
|
Front Panel Display Type |
6 digit LED
|
Front Panel Display Units |
K, °C, °F, V, Ω
|
Front Panel Display Update Rate |
2 rdg/s
|
Front Panel Temperature Display Resolution |
0.001° between 0‐99.999°, 0.01° between 100‐999.99°, 0.1° above 1000°
|
Front Panel Sensor Units Display Resolution |
100uV (for diode type), Sensor Dependent (for RTD type)
|
Front Panel Keys |
Menu, ˆ (Up Arrow), ˇ (Down Arrow), Enter
|
Front Panel Features |
View & Edit Configuration Settings, Instrument Reset
|
Serial Interface Format |
RS‐232C
|
Serial Interface Baud Rates |
9600, 19600, 57600, 115200
|
Serial Interface Functionality |
Graphical User Interface, LabView Drivers (Version 8.0 and above)
|
Serial Interface Connector |
DB9
|
Ethernet Interface Format |
10‐BaseT
|
Ethernet Interface Speed |
10 Mbps
|
Ethernet Interface Functionality |
Graphical User Interface, LabView Drivers , Web Based Monitoring Utility
|
Ethernet Interface Connector |
RJ‐45
|
Alarm Number |
2, High and Low
|
Alarm Settings |
High Setpoint, Low Setpoint, Dead band, Latching or Non‐latching, Alarm On/Off
|
Alarm Actuators |
Front Panel LEDs, relays
|
Number of Relays |
2
|
Contacts |
Normally Open (NO), Normally Closed (NC), and Common (C)
|
Relay Contact Rating |
24 VDC at 1 A
|
Relay Settings |
Manually off, manually on, follows alarms
|
Connector |
DB15
|
Analog Output Isolation |
Output is not isolated from chassis ground
|
Analog Output Update Rate |
2rdg/s
|
Voltage/Current Output |
RANGE:0‐10V, 4‐20mA
RESOLUTION:0.16 mV0.3 μA
ACCURACY:±1.25 mV±2.5 μA
MIN. LOAD RESISTANCE500 Ω
(short‐circuit protected) N/A
COMPLIANCE VOLTAGEN/A10 V
LOAD REGULATION:N/A10 V
|
Settings |
Scale voltage or current with temperature, native sensor units, or natural logarithm of sensor
units
|
Power Requirement |
Regulated 9‐24 VDC at 500 mA, Barrel Plug 5.5 mm OD x 2.5 mm ID x 9.9 mm L, Center Positive
|
Size |
1/8 DIN
|
Supported OS |
Windows 7.0, XP & Vista (32 bit)
|
Mounting |
Panel mount into 91 mm W x 44 mm H (3.6 x 1.7 in) cutout
|
Weight |
0.45 kg (1 lb)
|
Approval |
CE mark, RoHS compliant
|